sales@anandscientific.com Phone: +91-9899199946, 9899633313
plant growth

Plant Growth Chamber

With the solid double walled door, a full view inner plexi-glass door enables inspection of specimens inside the chamber without disturbing the process temperature.
The excellent quality of magnetic gum packing for the external door.
With forced air circulation at triple walled back, it maintains optimum uniformity of distribution of temperature.
High grade Polyurethane Foam (PUF) insulation between outer and inner chamber for minimal thermal loss.
Caster wheel mounted exterior cabinet is made of steel sheet duly powder coated.
Inner chamber and trays are made of stainless steel sheet (SS-304 grade).
Front double walled door is provided with lock and key arrangement.
Door operated illumination lamp is fitted inside the chamber for easy visibility.
Exterior illumination with fluorescent lamps/tubes accounts for artificial daylight conditions.
Finned tube evaporator for uniform and faster cooling effects.
Temperature range 5C to 60C 1C controlled with an accuracy of 1C.
When all lights are ON temperature range is 15C to 60C.
Safety thermostat to prevent overheating.
Fitted with 162 line (32 characters) alphanumeric backlit LCD Display with membrane keyboard with tactile feedback.
Data (temperature, humidity and light levels) storage upto 1000 readings.
Programmable time interval from 0-99 minutes.
Real Time Clock.
Audio/visual alarm for temperature humidity deviations from set values.
Five Levels light control (0%, 25%, 50%, 75% and 100%).
Program lock to avoid any changes in control parameters by unauthorized
people.
Watchdog timer
Temperature Sensor : PT-100
Programmable time intervals for temperature, humidity and light level control
CFC free eco-friendly compressor.
Humidity is controlled (5% above ambient) from 55% RH to 95%RH at cool temperature.
With high-performance performance 9 segments, it enables end-users to control nine different variation and repetition of temperature, illumination and humidity according to its condition.
Power Supply : 220-230 Volts AC (50/60 Hz frequency) & supplied with 3 Step Stability Chamber

WORKING CHAMBER SIZE (W x D x H )

VOLUME

505 mm

6.1 Cuft

570 mm

10 Cuft

650 mm

12 Cuft

700 mm

15 Cuft

775 mm

20 Cuft

825 mm

30 Cuft

900 mm

45 Cuft

Pre Installation Mandatory Requirement :
(Guarantee void in case of non compliance)

a) Distilled water or DM water continuous supply with tap for boiler input of 1/2 BSP at 3 feet height
b) Room temperature around machine preferably at 25C with air conditioning or a well ventilated room with exhaust fan. However surrounding temperature should not exceed 30C
c) Stabilized input voltage of 230V AC 20 Amp. Use of servo controlled stabilizer is
recommended
d) Water drain line with 1/2″ nozzle to be provided at floor level

GMP Model
Outer cabinet SS 304 grade dull matt finish
Inner chamber stainless steel mirror ?nish
With DQ. IQ. OQ documentation

Cu. Ft. WxDxH mm SS

4 112 455 x 410 x 610mm 2
6.1 171 505 X 415 X 830 2
l0 230 570 x 550 x 875 3
12 336 650 x sxo x 900 3
15 420 700 x 640 x 900 4

Optional Extras Accessories:
a) PLC based system PLC with HMI 3.7 touch screen grey monitor with 21 CFR Pan 11 features (OSLONG) software
b) Magnetic door lock feature
c) GSM system alarm sent to predefined mobile numbers.
d) Digital PID controller with printer interface to connect EPSON Dot matrix line printer. Print interval programmable can print date. time. Temperature.
e) PC interface for above controller with software on CD & hardware or software with 21 CFR part 11 features.
f) lmported RH sensor European make, Rotronic Switzerland for higher accuracy
g) Standby humidity system
h) Standby refrigeration system
i) Data logger: 4 point temp. 9 4 point humidity data logger with sensors placed at speci?c points in chamber with printer interface and memory of S000 readings per channel